Reported Mar 15, 2017 · Terminated
IMMULITE/ IMMULITE 1000 Progesterone; Catalog Number: LKPW1; Siemens Material Number: 10381128 Product Usage: For in…
Siemens Healthcare Diagnostics, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Healthcare Diagnostics, Inc · 333 Coney St · East Walpole, MA · 02032-1516
Is this the product you have?
- IMMULITE/ IMMULITE 1000 Progesterone
- Catalog NumberLKPW1
- Siemens Material Number10381128
- Product UsageFor in vitro diagnostic use with the IMMULITE¿2000 Systems Analyzers- for the quantitative measurement of progesterone in serum, as an aid in the diagnosis and treatment of disorders of the ovaries or placenta
- Lot codes and dates
- LKPW1 kit lot 0245 LKPW1 kit lot 0246 LKPW1 kit lot 0247 LKPW1 kit lot 0248 LKPW1 kit lot 0249
- Amount recalled
- 9076
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
- Nationwide
- PR
Why it was recalled
DHEA-S causes falsely elevated progesterone results.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.