Reported Mar 15, 2017 · Terminated

IMMULITE/ IMMULITE 1000 Progesterone; Catalog Number: LKPW1; Siemens Material Number: 10381128 Product Usage: For in…

Siemens Healthcare Diagnostics, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Healthcare Diagnostics, Inc · 333 Coney St · East Walpole, MA · 02032-1516

Is this the product you have?

  • IMMULITE/ IMMULITE 1000 Progesterone
  • Catalog NumberLKPW1
  • Siemens Material Number10381128
  • Product UsageFor in vitro diagnostic use with the IMMULITE¿2000 Systems Analyzers- for the quantitative measurement of progesterone in serum, as an aid in the diagnosis and treatment of disorders of the ovaries or placenta
Lot codes and dates
  • LKPW1 kit lot 0245 LKPW1 kit lot 0246 LKPW1 kit lot 0247 LKPW1 kit lot 0248 LKPW1 kit lot 0249
Amount recalled
9076

This recall covers 4 products

The agency filed one notice per product. You are reading one of them; the other 3 are below.

Where it was sold

  • Nationwide
  • PR

Why it was recalled

DHEA-S causes falsely elevated progesterone results.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jul 18, 2019 · Data as of Sep 19, 5:30 PM UTC