Reported Mar 11, 2020 · Terminated
Bard Biopsy EnCor Probe, Stereotactic/Ultrasound 7G, REF number ECP017G, packaged individually in sterile pouches, 5…
Bard Peripheral Vascular Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Bard Peripheral Vascular Inc · 1625 W 3rd St Bldg 1 · Tempe, AZ · 85281-2438
Is this the product you have?
- Bard Biopsy EnCor Probe, Stereotactic/Ultrasound 7G, REF number ECP017G, packaged individually in sterile pouches, 5 pouches/carton, Single Use, Rx, Sterile.
- Lot codes and dates
Lot numbers VTCT0268VTCT0269VTCT0270VTCT0271VTCU0340and VTCU0341
- UDI numbers
(01)00801741086281(17)200601(10)VTCT0268(01)00801741086281(17)200601(10)VTCT0269(01)00801741086281(17)200601(10)VTCT0270(01)00801741086281(17)200601(10)VTCT0271(01)00801741086281(17)200601(10)VTCU0340and (01)00801741086281(17)200601(10)VTCU0341
- Amount recalled
- 12,875 units
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
Why it was recalled
Lot numbers and products inadvertently not included in the scope of the previous recall. Original recall was conducted due to an increase in probe failures for leaks, suction issues, and failure to obtain samples.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Bard Peripheral Vascular Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.