Reported Mar 18, 2026 · Ongoing

Owner's Booklets and Instructions for Use that are used with the following blood glucose measurement devices.…

Trividia Health, Inc.

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Trividia Health, Inc. · 2400 Nw 55th Ct · Fort Lauderdale, FL · 33309-2672

Is this the product you have?

  • Owner's Booklets and Instructions for Use that are used with the following blood glucose measurement devices
  • Description/ItemTRUE METRIX PRO Meter Only/RE4H01P-40, TRUE METRIX PRO Starter Kit/RE4H01P-43, PROCURE TRUE METRIX PRO Meter Only/RE4212P-00, Henry Schein TRUE METRIX PRO Meter Only/RE4099P-40, Henry Schein TRUE METRIX PRO Kit/RE4099P-43, McKESSON TRUE METRIX PRO Meter Only/RE4051P-00, Moore Medical TRUE METRIX PRO Meter Only/RE4068P-01
Lot codes and dates
  • Device Item/UDI-DI
    • RE4H01P-40/21292006099
    • RE4H01P-43/21292006105
    • RE4212P-00/10840330705681
    • RE4099P-40/21292014919
    • RE4099P-43/21292008673
    • RE4051P-00/20612479197217
    • RE4068P-01/607415261097. Owner's Booklet
  • or IFU Description/Part and RevisionTRUE METRIX PRO Owners Booklet/RE4TVHP03 Rev 56 and prior, McKesson Med Surg TRUE METRIX PRO Owners Booklet/ RE4SUNP03 Rev 56 and prior
  • Device serial numbersPrefix "TP" followed by a seven-digit numeric sequence
Amount recalled
1,001,534

This recall covers 4 products

The agency filed one notice per product. You are reading one of them; the other 3 are below.

Where it was sold

  • Nationwide
  • PR

Why it was recalled

Stop using this product now

The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Trividia Health, Inc. recalls