Reported Mar 18, 2026 · Ongoing

DxC 700 AU, REF: B86444, B86446

Beckman Coulter Mishima K.K.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Beckman Coulter Mishima K.K. · 454-32 · Sunto-Gun, N/A · N/A

Is this the product you have?

  • DxC 700 AU
  • REF
    • B86444
    • B86446
Lot codes and dates
  • REF/UDI-DI/Serial NumbersNewer Instrument Installed with PC preinstalled
  • with Microsoft SQL Server 2022 ExpressB86444/14987666542545/2025060063, 2025060073, 2025060074, 2025060075, 2025060079, 2025060085, 2025060086, 2025060087, 2025060091, 2025060098, 2025060099, 2025060124, 2025060125, 2025060126, 2025060127, 2025060128, 2025060129, 2025070139, 2025070156, 2025070159, 2025070160, 2025070161, 2025070164, 2025070167, 2025070168, 2025070169, 2025070178, 2025070180, 2025070181, 2025070182, 2025070184, 2025070186, 2025070191, 2025070192, 2025070193, 2025070194, 2025070204, 2025080208, 2025080217, 2025080219, 2025080236, 2025080237, 2025080238, 2025080241, 2025080242, 2025080246, 2025080247, 2025080250, 2025090266, 2025090267, 2025090270, 2025090271, 2025090273, 2025090274, 2025090275, 2025090281, 2025090283, 2025090285, 2025090286, 2025090287, 2025090289, 2025090290, 2025090291, 2025090292, 2025090293, 2025090294, 2025090295, 2025090296, 2025090297, 2025090298, 2025090299, 2025090301, 2025090302, 2025090303, 2025090304, 2025090306, 2025100370, 2025100371, 2025100372, 2025100373, 2025100421, 2025100422, 2025100423, 2025100424, 2025100425, 2025100426, 2025100427, 2025110428, 2025110429, 2025110437, 2025110442, 2025110445, 2025110452, 2025110454, 2025110456, 2025110457, 2025110458, 2025110462, 2025120464, 2025120468, 2025120469, 2025120471, 2025120475, 2025120477, 2025120478, 2025120479, 2025120480, 2025120482, 2025120483, 2025120484, 2025120485, 2025120486, 2025120487, 2025120488, 2025120489, 2025120490, 2025120492, 2025120493, 2025120494, 2025120495, 2025120496. B86446/14987666542514/2025090277, 2025090284. Older Instrument Replaced with PC preinstalled
  • with Microsoft SQL Server 2022 Express
    • B86444/14987666542545/2017050064
    • 2017100174
    • 2018030312
    • 2018040352
    • 2018100561
    • 2018110599
    • 2019010712
    • 2019010713
    • 2019111170
    • 2021021852
    • 2021102318
    • 2021122458
    • 2022042631
    • 2022052705
    • 2022062775
    • 2023063701
    • 2024064408
    • 2024064419
    • 2024064440
Amount recalled
142

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Beckman Coulter Mishima K.K. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:16 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.