Reported Oct 7, 2026 · Ongoing · Updated Oct 13, 2026

Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 /…

Beckman Coulter Mishima K.K.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Beckman Coulter Mishima K.K. · 454-32 · Sunto-Gun · N/A

Is this the product you have?

  • Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY
  • ANALYZER AU480 REF / UDI-DI codeB96692 / 14987666542965 B96693 / 14987666542972 B12183 / 14987666540671 The AU480 Chemistry Analyzer measure analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material, and other accessories. This system is for in vitro diagnostic use only
Lot codes and dates

5 of 5 products

Products covered by this recall
Model / numberBarcodesProduct as published
—Item Numbers / UDI-DI codes
B9669214987666542965UDI-DI code: 14987666542965
B9669314987666542972UDI-DI code: 14987666542972
B1218314987666540671UDI-DI code: 14987666540671
N3151400UDI-DI code: Due to the age of the device and the fact that it has exceeded its storage period, the UDI number is not available.)
Amount recalled
492 units

Where it was sold

Why it was recalled

Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Beckman Coulter Mishima K.K. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Nov 4, 12:12 AM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.