Reported Aug 28, 2024 · Ongoing

DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in…

Beckman Coulter Mishima K.K.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Beckman Coulter Mishima K.K. · 454-32 · Sunto-Gun · N/A

Is this the product you have?

  • DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.
Lot codes and dates
  • Catalog NumberC63520
  • UDI-DI code14987666545065
  • Serial Numbers
    • 2023070010
    • 2023060005
    • 2023090015
    • 2023070009
    • 2023080011
    • 2023080012
    • 2023100019
    • 2023100018
    • 2023100020
    • 2024010037
    • 2024010036
    • 2023110022
    • 2024010038
    • 2023090016
    • 2024020040
    • 2024020041
    • 2024020039
    • 2024020043
    • 2024020042
    • 2023110023
    • 2024020045
    • 2024020046
    • 2024010024
    • 2024010025
Amount recalled
24 analyzers

Where it was sold

Why it was recalled

Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will be reported. This issue may cause a delay of results.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Beckman Coulter Mishima K.K. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:55 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.