Reported May 9, 2018 · Terminated
MultiDim Eleva 708032 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography…
Philips Medical Systems Nederlands
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Medical Systems Nederlands · Veenpluis 4-6 · Best, N/A · N/A
Is this the product you have?
- MultiDim Eleva 708032 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.
- Lot codes and dates
- System Serial # 32 177 259 239 264 258 52 67 111 257 94 142 121 191 143 260 265 74 207 101 182 174 124 263 4 172 254 151 246 64 183 76 152 188 261 242 102 34 147 241 134 178 245 176 175 38 162 107 220 65 194 69 247 13 215 78 112 29 139 98 79 158 61 33 106 35 50 16 80 205 161 11 204 37 12 189 9 232 15 228 155 47 171 213 209 218 96 9071 141 14 36 48 224 25 105 57 216 138 24 28 211 160 192 43 44 157 154 41 219 8 68 129 75 221 7 10 63 146 45 26 125 193 210 1 60 3 132 81 22 23 208 5 131 135 88 226 6 156 49 91 92 93 201 250 251 72 95 148 39 31 2 30 51
212217181163140114115108
83 159 53 137 54 116 118 119 167 168 169 97100255252256109196164170227173
- System Serial # 32 177 259 239 264 258 52 67 111 257 94 142 121 191 143 260 265 74 207 101 182 174 124 263 4 172 254 151 246 64 183 76 152 188 261 242 102 34 147 241 134 178 245 176 175 38 162 107 220 65 194 69 247 13 215 78 112 29 139 98 79 158 61 33 106 35 50 16 80 205 161 11 204 37 12 189 9 232 15 228 155 47 171 213 209 218 96 90
- Amount recalled
- 184
This recall covers 7 products
The agency filed one notice per product. You are reading one of them; the other 6 are below.
Where it was sold
- Nationwide
Why it was recalled
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips Medical Systems Nederlands recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.