Reported Jun 3, 2015 · Terminated

SOMATOM Emotion 6; the intended use of computed tomography is to produce cross-sectional images of the body by computer…

Siemens Medical Solutions USA, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Medical Solutions USA, Inc · 51 Valley Stream Pkwy · Malvern, PA · 19355-1406

Is this the product you have?

  • SOMATOM Emotion 6; the intended use of computed tomography is to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.
Lot codes and dates
  • SOMATOM Emotion 6 (2007) - 10165888
  • with serial numbers29338 27232 27230 27231
Amount recalled
75

This recall covers 7 products

The agency filed one notice per product. You are reading one of them; the other 6 are below.

Where it was sold

  • Nationwide
  • PR

Why it was recalled

Software bugs in VC20B SP0a or SP1 software versions may cause issues that could make it necessary to rescan patients. Syngo Main UI may crash if the Patient Browser is scrolled with arrow keys of the keyboard. Tomo images may freeze on the screen when zoomed in or out under the mode of CAREVision. Sporadic displaying error in WorkStream4D application. DB may lock under heavy and multitasks.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Medical Solutions USA, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Sep 1, 2016 · Data as of Sep 19, 6:04 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.