Reported Jun 3, 2015 · Terminated

SIEMENS Luminos Agile Max; a universal imaging system for radiographic and fluoroscopic studies. Using either film…

Siemens Medical Solutions USA, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Medical Solutions USA, Inc · 51 Valley Stream Pkwy · Malvern, PA · 19355-1406

Is this the product you have?

  • SIEMENS Luminos Agile Max; a universal imaging system for radiographic and fluoroscopic studies. Using either film cassettes or a digital mobile flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. Luminos Agile is applicable to emergency treatment on an outpatient basis, as well as for bedside examinations.
Lot codes and dates
  • Model Number
  • of device 10762472 with serial numbers
    • 61034
    • 61052
    • 61045
    • 61013
    • 61054
    • 61056
    • 61021
    • 61050
    • 61053
    • 61032
    • 61015
    • 61046
    • 61028
    • 61040
    • 61011
    • 61029
    • 61048
    • 61060
    • 61042
    • 61035
    • 61014
    • 61025
    • 61043
    • 61036
    • 61038
    • 61022
    • 61037
    • 61049
    • 61044
    • 61062
    • 61030
    • 61018
    • 61027
    • 61061
    • 61033
    • 61039
    • 61047
    • 61031
    • 61041
    • 61055
    • 61019
    • 61057
    • 61059
    • 61017
    • 61051
Amount recalled
133 total

This recall covers 7 products

The agency filed one notice per product. You are reading one of them; the other 6 are below.

Where it was sold

  • Nationwide

Why it was recalled

Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Medical Solutions USA, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Mar 20, 2017 · Data as of Sep 19, 6:16 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.