Reported Jun 11, 2014 · Terminated
IntelliSpace Portal (ISP), (ISP DX/HX/EX, Model number 881001; ISP IX, Model number 881030; ISP LX SPECT, Model number…
Philips Medical Systems (Cleveland) Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Medical Systems (Cleveland) Inc · 595 Miner Rd · Cleveland, OH · 44143-2131
Is this the product you have?
- IntelliSpace Portal (ISP), (ISP DX/HX/EX, Model number 881001; ISP IX, Model number 881030; ISP LX SPECT, Model number 881035), Philips Medical Systems. A medical software system for diagnostic imaging data viewing.
- Lot codes and dates
4 of 4 products
Products covered by this recall Model / number Product as published — ISP DX/HX/EX 881001Serial numbers 35033, 85298, 80861, 81017, 81021, 81191, 82085, 82124, 82144, 82148, 82242, 85029, 85030, 85033, 85164, 85166, 85183, 85247, 85302, 85324, 85337, 85353, 85357, 85359, 85397, 85401, 85402, 87004, 87014, 87158, 87176, 87177, 87182, 87200, 92003, 95043, 95044, 95045, 95046, 95047, 95048, 95049, 95050, 95051, 95052, 95054, 850001, 870030, 870071, 870073, 870076, 920007, 920008, 950006, 950023, 950031, 950039, 950041, 950045, 950052, 950071, 950084, 950085, 950087, 950106, 950119, 950120, 950121, 950129, 950137, 950143 and 950146. ISP IX 881030Serial numbers 40024, 40021, 40031, 40036, 40044, 40061, 40082, 40088, 40096, 40145, 40146, 40148, 40150, 40155, 40158, 40160, 40169 and 600003. ISP LX SPECT 881035Serial numbers 40001, 40003, 40033, 40084, 40087, 40093, 40133, 40134 and 40141. - Amount recalled
- 157 units
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
- Nationwide
Why it was recalled
During SPECT reconstructions using Attenuation Correction and Scatter Correction, no Scatter Correction is being applied in the AutoSPECT Pro application on IntelliSpace Portal. In addition, resolution recovery is not applied correctly in SPECT reconstructions using Astonish in AutoSPECT Pro on Intellispace Portal.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips Medical Systems (Cleveland) Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.