Reported Apr 26, 2017 · Terminated
Allura Xper Release 8.2 (Laird chiller is always installed); Allura Xper FD10 722026; Allura Xper FD10/10 722027;…
Philips Electronics North America Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Electronics North America Corporation · 3000 Minuteman Rd · Andover, MA · 01810-1032
Is this the product you have?
9 of 9 products
| # | Product as published |
|---|---|
| 1 | Allura Xper Release 8.2 (Laird chiller is always installed) |
| 2 | Allura Xper FD10 722026 |
| 3 | Allura Xper FD10/10 722027 |
| 4 | Allura Xper FD20/10 biplane 722029 |
| 5 | Allura Xper FD10 OR Table 722033 |
| 6 | Allura Xper FD10/10 OR Table 722034 |
| 7 | Allura Xper FD20/10 OR Table 722036 |
| 8 | Allure Xper FD20/20 722038 |
| 9 | Allura Xper FD20/20 biplane OR Table 722039 |
- Lot codes and dates
- Devices with Laird Chiller.
- Amount recalled
- 6992 (total for all devices)
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Where it was sold
- Nationwide
Why it was recalled
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips Electronics North America Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.