Reported Apr 26, 2017 · Terminated
UNIQ: UNIQ FD10 722026; UNIQ FD10/10 722027; UNIQ FD20/10 biplane 722029; UNIQ FD10 OR Table 722033; UNIQ FD10/10 OR…
Philips Electronics North America Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Electronics North America Corporation · 3000 Minuteman Rd · Andover, MA · 01810-1032
Is this the product you have?
- UNIQ
UNIQ FD10 722026UNIQ FD10/10 722027UNIQ FD20/10 biplane 722029UNIQ FD10 OR Table 722033UNIQ FD10/10 OR Table 722034UNIQ FD20/10 OR Table 722036UNIQ FD20/20 722038UNIQ FD20/20 biplane OR Table 722039
- Lot codes and dates
- Devices with a Laird Chiller.
- Amount recalled
- 6992 (total for all devices)
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Where it was sold
- Nationwide
Why it was recalled
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips Electronics North America Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.