Reported Jul 16, 2014 · Terminated
The STERIS 4085 General Surgical Table Product Usage: is an electro-hydraulically operated surgical table designed to…
Steris Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Steris Corporation · 5960 Heisley Rd · Mentor, OH · 44060-1834
Is this the product you have?
- The STERIS 4085
- General Surgical Table Product Usageis an electro-hydraulically operated surgical table designed to support all general surgical procedures
- Lot codes and dates
- All Model Numbers
- Amount recalled
- 3689
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
STERIS engineering analysis has determined the D1 pressure switch in the hydraulic column manifold does not adequately detect pressure level. If one or more of the table floor lock feet become unlocked due to the loss of hydraulic pressure, there is a potential for the table hand control to incorrectly indicate that the floor lock feet are locked.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Steris Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.