Reported May 24, 2017 · Terminated
Proton therapy system -Proteus 235 aka Proteus Plus. A medical device designed to produce and deliver a proton beam for…
Ion Beam Applications S.A.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Ion Beam Applications S.A. · Chemin du Cyclotron, 3 · Louvain La Neuve · N/A
Is this the product you have?
- Proton therapy system -Proteus 235 aka Proteus Plus. A medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
- Lot codes and dates
- Serial numbers
PAT.003PAT.006
- Serial numbers
- Amount recalled
- 2
Where it was sold
Why it was recalled
A PBS (Pencil Beam Scanning) irradiation may pause for different reasons. After a pause, irradiation may restart from the beginning instead of recalculating the field based on the already delivered dose.
This product could cause temporary or reversible health problems — stop using it.