Reported Jun 10, 2020 · Terminated

Xhibit Central, Model No. 96102 - Product Usage: intended use is to provide clinicians with central monitoring of…

Spacelabs Healthcare, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Spacelabs Healthcare, Inc. · 35301 Se Center St · Snoqualmie, WA · 98065-9216

Is this the product you have?

  • Xhibit Central, Model No. 96102 -
  • Product Usageintended use is to provide clinicians with central monitoring of adult, pediatric and neonatal patient data of patients connected to networked Spacelabs Healthcare patient monitors and telemetry transmitters
Lot codes and dates
  • Model No. 96102 Xhibit Central
  • Software Versions
    • 1.3.1
    • 1.3.2
    • 1.3.3 and 1.3.4
  • UDI10841522100345
Amount recalled
2197 units

Where it was sold

  • Nationwide
  • PR

Why it was recalled

The firm received reports of loss of audio alarm after a power failure or cable disconnection. The loss of audio for a telemetry patient could delay the recognition of an alarm condition.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Spacelabs Healthcare, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 18, 2024 · Data as of Sep 19, 5:32 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.