Reported Jun 10, 2020 · Terminated
Xhibit Central, Model No. 96102 - Product Usage: intended use is to provide clinicians with central monitoring of…
Spacelabs Healthcare, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Spacelabs Healthcare, Inc. · 35301 Se Center St · Snoqualmie, WA · 98065-9216
Is this the product you have?
- Xhibit Central, Model No. 96102 -
- Product Usageintended use is to provide clinicians with central monitoring of adult, pediatric and neonatal patient data of patients connected to networked Spacelabs Healthcare patient monitors and telemetry transmitters
- Lot codes and dates
- Model No. 96102 Xhibit Central
- Software Versions
1.3.11.3.21.3.3 and 1.3.4
- UDI10841522100345
- Amount recalled
- 2197 units
Where it was sold
- Nationwide
- PR
Why it was recalled
The firm received reports of loss of audio alarm after a power failure or cable disconnection. The loss of audio for a telemetry patient could delay the recognition of an alarm condition.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Spacelabs Healthcare, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.