Reported Oct 9, 2019 · Terminated

Accu-Chek Performa II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to…

Roche Diabetes Care, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Roche Diabetes Care, Inc. · 9115 Hague Rd · Indianapolis, IN · 46256-1025

Is this the product you have?

  • Accu-Chek
  • Performa II Meter - Product Usagea. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. b. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Accu-Chek Aviva Plus Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly). c. The Accu-Chek Aviva Plus Test Strips are for use with the Accu-Chek Aviva Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm
Lot codes and dates
  • Serial numbers less than xxx25525056
  • Updated serial numbers on 10/12/2019Performa
  • IIBetween XXX20000000
  • and XXX25525056 Updated 10/18/2019Performa
  • IIfrom XXX20000000 to XXX25525056
Amount recalled
N/A

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

  • Nationwide

Why it was recalled

Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on).

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Roche Diabetes Care, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Oct 30, 2020 · Data as of Sep 19, 6:59 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.