Reported Dec 6, 2017 · Terminated

Dimension Vista Protein 1 Calibrator, PROT 1 CAL, Siemens Material Number (SMN) 10465664 (US) and 10469516 (OUS),…

Siemens Healthcare Diagnostics, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Healthcare Diagnostics, Inc. · 500 Gbc Dr Ms 514 · Newark, DE · 19702-2466

Is this the product you have?

  • Dimension Vista Protein 1 Calibrator, PROT 1 CAL, Siemens Material Number (SMN) 10465664 (US) and 10469516 (OUS), Product Code KC710U (US) and KC710 (OUS); IVD
Lot codes and dates
  • Lot numbers6JC001, 6JC001C, 6JC001G (exp. date 2018-01-11); 6MC001, 6MC001C (exp. date 2018-04-29); 6CC001A, 6CC001B, 6CC001F (exp. date 2017-07-14); 6FC001A, 6FC001C, 6FC001E (exp. 2017-10-27); --- Lot # (UDI): 6CC001A (008427680244436CC001A20170714), 6CC001B (008427680244436CC001B20170714), 6CC001F (008427680244436CC001F20170714), 6FC001A (008427680244436FC001A20171027), 6FC001C (008427680244436FC001C20171027), 6FC001E (008427680244436FC001E20171027), 6JC001 (008427680244436JC00120180111), 6JC001C (008427680244436JC001C20180111), 6JC001G (008427680244436JC001G20180111), 6MC001 (008427680244436MC00120180429), 6MC001C (008427680244436MC001C20180429)
Amount recalled
N/A

This recall covers 4 products

The agency filed one notice per product. You are reading one of them; the other 3 are below.

Where it was sold

  • Nationwide

Why it was recalled

Complaint of high amount of ceruloplasmin patient results below the reference range (< 20 mg/dL [0.2 g/L]. PROT1 CAL and corresponding PROT1 CON L, M, and H are inaccurately assigned for ceruloplasmin (CER).

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Mar 30, 2018 · Data as of Sep 19, 4:57 PM UTC