Reported Dec 6, 2017 · Terminated

Dimension Vista Protein 1 Control M, PROT 1 CON M, Siemens Material Number (SMN) 10445920, Product Code KC716; IVD

Siemens Healthcare Diagnostics, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Healthcare Diagnostics, Inc. · 500 Gbc Dr Ms 514 · Newark, DE · 19702-2466

Is this the product you have?

    • Dimension Vista Protein 1 Control M
    • PROT 1 CON M
    • Siemens Material Number (SMN) 10445920
    • Product Code KC716
    • IVD
Lot codes and dates
  • Lot numbers6BQM01, 6BQM01A, 6BQM01B, 6BQM01C,6BQM01D, 6BQM01E (exp. date 2017-06-17); 6FQM01, 6FQM01A (exp. date 2017-11-03) --- Lot # (UDI): 6BQM01 (008427680063336JQL01C20180131), 6BQM01A (008427680063406BQM01A20170617), 6BQM01B (008427680063406BQM01B20170617), 6BQM01C (008427680063406BQM01C20170617), 6BQM01D( 008427680063406BQM01D20170617), 6BQM01E (008427680063406BQM01E20170617), 6FQM01(008427680063406FQM0120171103), 6FQM01A (008427680063406FQM01A20171103)
Amount recalled
N/A

This recall covers 4 products

The agency filed one notice per product. You are reading one of them; the other 3 are below.

Where it was sold

  • Nationwide

Why it was recalled

Complaint of high amount of ceruloplasmin patient results below the reference range (< 20 mg/dL [0.2 g/L]. PROT1 CAL and corresponding PROT1 CON L, M, and H are inaccurately assigned for ceruloplasmin (CER).

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Mar 30, 2018 · Data as of Sep 19, 6:00 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.