Reported Dec 6, 2017 · Terminated

Dimension Vista Protein 1 Control H, PROT 1 CON H, Siemens Material Number (SMN) 10445921, Product Code KC717; IVD.

Siemens Healthcare Diagnostics, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Healthcare Diagnostics, Inc. · 500 Gbc Dr Ms 514 · Newark, DE · 19702-2466

Is this the product you have?

    • Dimension Vista Protein 1 Control H
    • PROT 1 CON H
    • Siemens Material Number (SMN) 10445921
    • Product Code KC717
    • IVD
Lot codes and dates
  • Lot numbers6EQH01, 6EQH01A, 6EQH01B, 6EQH01C (exp. date 2017-09-17); 6LQH01A, 6LQH01B, 6LQH01C (exp. date 2018-04-04). --- Lot # (UDI): 6EQH01 (008427680063576EQH0120170917), 6EQH01A (008427680063576EQH01A2017091), 6EQH01B(008427680063576EQH01B20170917), 6EQH01C(008427680063576EQH01C20170917), 6LQH01A (008427680063576LQH01A20180404), 6LQH01B (008427680063576LQH01B20180404), 6LQH01C (008427680063576LQH01C20180404)
Amount recalled
N/A

This recall covers 4 products

The agency filed one notice per product. You are reading one of them; the other 3 are below.

Where it was sold

  • Nationwide

Why it was recalled

Complaint of high amount of ceruloplasmin patient results below the reference range (< 20 mg/dL [0.2 g/L]. PROT1 CAL and corresponding PROT1 CON L, M, and H are inaccurately assigned for ceruloplasmin (CER).

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Mar 30, 2018 · Data as of Sep 19, 4:46 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.