Reported Sep 27, 2023 · Ongoing
Hamilton-T1, REF: MSP161339/02, ESM Shielding Set, Spare Part for ventilator
Hamilton Medical, Inc.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Hamilton Medical, Inc. · 4655 Aircenter Cir · Reno, NV · 89502-5948
Is this the product you have?
- Hamilton-T1
- REFMSP161339/02, ESM Shielding Set, Spare Part for ventilator
- Lot codes and dates
- UDI07630002800761 /
- Affected Serial Numbers
405019340502104050194 and 4016669
- Amount recalled
- 4 shielding set
This recall covers 6 products
The agency filed one notice per product. You are reading one of them; the other 5 are below.
Where it was sold
- Nationwide
- PR
Why it was recalled
Stop using this product now
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Hamilton Medical, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.