Reported Oct 1, 2025 · Ongoing
Philips CT systems labeled as the following: 1. Brilliance 16 Power, Model Number: 728240. 2. Brilliance CT 16 Slice,…
Philips North America LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips North America Llc · 222 Jacobs St · Cambridge, MA · 02141-2289
Is this the product you have?
- Philips
- CT systems labeled as the following1. Brilliance 16 Power
- Model Number728240. 2. Brilliance CT 16 Slice
- Model Number728246. 3. Brilliance CT 6 Slice
- Model Number728256
- Lot codes and dates
3 of 3 products
Products covered by this recall Model / number Product as published 728240UDI-DI: N/A; Serial Number: 6022. 728246UDI-DI: N/A; Serial Numbers: 5029, 5698, 5834, 6272, 6313, 5307, 5305, 50129, 5286, 50017, 5388, 5424, 5422, 5515, 5497, 5584, 5668, 5624, 5771, 5814, 5872, 5907, 5900, 5928, 5869, 5939, 5944, 5993, 6137, 5411, 6160, 6153, 5504, 6182, 6199, 6209, 6219, 6271, 6302, 6330, 6357, 6358, 6369, 6405, 6343, 6410, 6558, 6490, 6635, 6532, 6590, 6547, 6669, 6691, 6563, 6756, 6755, 6738, 3578, 6776, 260046, 260074, 6752, 50219, 50028, 5412, 5839. 728256UDI-DI: N/A; Serial Numbers: 30117, 3290, 3347, 3664, 3499. - Amount recalled
- 73 units
This recall covers 10 products
The agency filed one notice per product. You are reading one of them; the other 9 are below.
Where it was sold
- Nationwide
Why it was recalled
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips North America LLC recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.