Reported Oct 1, 2025 · Ongoing
Philips CT systems labeled as the following: 1. IQon Spectral CT, Model Number: 728332.
Philips North America LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips North America Llc · 222 Jacobs St · Cambridge, MA · 02141-2289
Is this the product you have?
- Philips
- CT systems labeled as the following1. IQon Spectral CT
- Model Number728332
- Lot codes and dates
3 of 3 products
Products covered by this recall Model / number Barcodes Product as published 728332— — 00884838059542UDI-DI: 00884838059542 — Serial Numbers: 860088, 860023, 860077, 860086, 60019, 860152, 860215, 60031, 60123, 60129, 60152, 860011, 860049, 860019, 860008, 860030, 860025, 860009, 860022, 860013, 860024, 860020, 860036, 860041, 860046, 860055, 860058, 860054, 860073, 860065, 860070, 860072, 860066, 860103, 860076, 860085, 860095, 860102, 860101, 860104, 860175, 860136, 860117, 860120, 860121, 860124, 860125, 860191, 860131, 860150, 860164, 60041, 860207, 60003, 860157, 860167, 860158, 860196, 60012, 860176, 860185, 860168, 860179, 860194, 60010, 60013, 60007, 60001, 395003, 860209, 860208, 60036, 60030, 60027, 60054, 60055, 60053, 60058, 60079, 395005, 60145. - Amount recalled
- 81 units
This recall covers 10 products
The agency filed one notice per product. You are reading one of them; the other 9 are below.
Where it was sold
- Nationwide
Why it was recalled
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips North America LLC recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.