Reported Oct 1, 2025 · Ongoing
Philips CT systems labeled as the following: 1. Ingenuity CT Family Brazil, Model Number: 728325.
Philips North America LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips North America Llc · 222 Jacobs St · Cambridge, MA · 02141-2289
Is this the product you have?
- Philips
- CT systems labeled as the following1. Ingenuity CT Family Brazil
- Model Number728325
- Lot codes and dates
- 1
- Model Number728325
- UDI-DIN/A
- Serial Numbers
338014338028338019
- Amount recalled
- 3 units
This recall covers 10 products
The agency filed one notice per product. You are reading one of them; the other 9 are below.
Where it was sold
- Nationwide
Why it was recalled
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
This product could cause temporary or reversible health problems — stop using it.