Reported Oct 1, 2025 · Ongoing
Philips CT systems labeled as the following: 1. CT 6000, Model Number: 728307.
Philips North America LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips North America Llc · 222 Jacobs St · Cambridge, MA · 02141-2289
Is this the product you have?
- Philips
- CT systems labeled as the following1. CT 6000
- Model Number728307
- Lot codes and dates
- 1
- Model Number728307
- UDI-DI00884838104600
- Serial Numbers
871508716987118871688716687164
- Amount recalled
- 6 units
This recall covers 10 products
The agency filed one notice per product. You are reading one of them; the other 9 are below.
Where it was sold
- Nationwide
Why it was recalled
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips North America LLC recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.