Reported Oct 8, 2025 · Ongoing
MAGNETOM Connectom.X. Model Number: 11371480.
Siemens Medical Solutions USA, Inc.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355-1418
Is this the product you have?
- MAGNETOM Connectom.X
- Model Number11371480
- Lot codes and dates
- Model Number11371480
- UDI NumbersN/A
- Serial Numbers237551
- Amount recalled
- 1 unit
This recall covers 15 products
The agency filed one notice per product. You are reading one of them; the other 14 are below.
2 more are in the archive.
Where it was sold
- Nationwide
- GA
Why it was recalled
Stop using this product now
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Siemens Medical Solutions USA, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.