Reported Oct 8, 2025 · Ongoing
MAGNETOM Verio Dot. Model Number: 10684333.
Siemens Medical Solutions USA, Inc.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355-1418
Is this the product you have?
- MAGNETOM Verio Dot
- Model Number10684333
- Lot codes and dates
- Model Number10684333
- UDI Numbers
(01)04056869006673(21)60060(01)04056869006673(21)160040(01)04056869006673(21)160028(01)04056869006673(21)160021(01)04056869006673(21)60108(01)04056869006673(21)60121(01)04056869006673(21)160019(01)04056869006673(21)60091(01)04056869006673(21)160020(01)04056869006673(21)160043(01)04056869006673(21)60062(01)04056869006673(21)60061(01)04056869006673(21)160029(01)04056869006673(21)160038(01)04056869006673(21)60086(01)04056869006673(21)60126(01)04056869006673(21)60047(01)04056869006673(21)160036(01)04056869006673(21)160003(01)04056869006673(21)160042(01)04056869006673(21)160024(01)04056869006673(21)60092(01)04056869006673(21)160035(01)04056869006673(21)60102(01)04056869006673(21)160039(01)04056869006673(21)60112(01)04056869006673(21)160030(01)04056869006673(21)60099(01)04056869006673(21)160007(01)04056869006673(21)160001(01)04056869006673(21)160006(01)04056869006673(21)160009(01)04056869006673(21)160033(01)04056869006673(21)60053(01)04056869006673(21)60070(01)04056869006673(21)60101(01)04056869006673(21)60125(01)04056869006673(21)60085(01)04056869006673(21)60113(01)04056869006673(21)160027(01)04056869006673(21)160012(01)04056869006673(21)160016(01)04056869006673(21)160032(01)04056869006673(21)160013(01)04056869006673(21)160014(01)04056869006673(21)160018(01)04056869006673(21)160031(01)04056869006673(21)160011(01)04056869006673(21)160015(01)04056869006673(21)60054(01)04056869006673(21)160037(01)04056869006673(21)160034(01)04056869006673(21)60063(01)04056869006673(21)60079(01)04056869006673(21)160008(01)04056869006673(21)60110(01)04056869006673(21)60124(01)04056869006673(21)160041
- Serial Numbers
600601600401600281600216010860121160019600911600201600436006260061160029160038600866012660047160036160003160042160024600921600356010216003960112160030600991600071600011600061600091600336005360070601016012560085601131600271600121600161600321600131600141600181600311600111600156005416003716003460063600791600086011060124160041. Additional
- serial numbers added 2/25/202660113
- Amount recalled
- 59 units
This recall covers 15 products
The agency filed one notice per product. You are reading one of them; the other 14 are below.
2 more are in the archive.
Where it was sold
- Nationwide
- GA
Why it was recalled
Stop using this product now
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Siemens Medical Solutions USA, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.