Reported Jan 22, 2025 · Ongoing

GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, Emission

Ge Medical Systems Israel, Functional Imaging

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING · 4, Hayozma St · Tirat Carmel · N/A

Is this the product you have?

  • GE Healthcare Discovery NM/CT 670 CZT
  • Model/Catalog Numbers1) H3906AB; System, Tomography, Computed, Emission
Lot codes and dates

8 of 8 products

Products covered by this recall
#BarcodesProduct as published
1840682121446UDI/DI 840682121446, System ID UX2305NU03, Serial Number CZTV29005
2840682121446UDI/DI 840682121446, System ID M4166103, Serial Number CZTV29006
3840682121446UDI/DI 840682121446, System ID 00243NUC22, Serial Number CZTW29008
4840682121446UDI/DI 840682121446, System ID 00222NUC07, Serial Number CZTW29009
5UDI/DI N/A, System ID 604875D670B, Serial Number CZTW29007
6UDI/DI N/A, System ID NT6200, Serial Number AD7W38007
7UDI/DI N/A, System ID NT6201, Serial Number AD7X38017
8UDI/DI N/A, System ID NT6100, Serial Number AP7W37001.
Amount recalled
8 units

This recall covers 13 products

The agency filed one notice per product. You are reading one of them; the other 12 are below.

Why it was recalled

GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Ge Medical Systems Israel, Functional Imaging recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:36 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.