Reported Jan 22, 2025 · Ongoing

GE Healthcare Brivo NM 615, Model/Catalog Numbers: 1) 5376204-15, 2) H2401MA, 3) H3100TA; System, Tomography, Computed,…

Ge Medical Systems Israel, Functional Imaging

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING · 4, Hayozma St · Tirat Carmel · N/A

Is this the product you have?

  • GE Healthcare Brivo NM 615
  • Model/Catalog Numbers1) 5376204-15, 2) H2401MA, 3) H3100TA; System, Tomography, Computed, Emission
Lot codes and dates

113 of 113 products

Products covered by this recall
#BarcodesProduct as published
100840682145503UDI/DI 00840682145503, System ID 502629NMGD, Serial Number NT2UG2100002BH
200840682117869UDI/DI 00840682117869, System ID 509455BRIVO, Serial Number NT24G1700010BH, UDI/DI N/A, System ID 248544D630, Serial Number NT2QG1900003BH\NT2UG2100001BH. 2) H2401MA: UDI/DI Not Available, System ID 501552D615, Serial Number NT24G1600023
3100035454149100035454149UDI/DI Not Available, System ID 100035454149, Serial Number T2CG15015/100035454149
4UDI/DI Not Available, System ID 337643NM, Serial Number NT24G1600021
5UDI/DI Not Available, System ID 718963B615, Serial Number NT24G1700009
6UDI/DI Not Available, System ID Not Available, Serial Number NT24G1600020
7UDI/DI Not Available, System ID 509455BRIVO, Serial Number NT24G1700010
8UDI/DI Not Available, System ID 509455BRIVO, Serial Number NT24G1800001
900840682117869UDI/DI (01)00840682117869(11)17060021NT24G1700003BH, System ID NM6583AR, Serial Number NT24G1700003
1000840682117869UDI/DI (01)00840682117869(11)17090021NT24G1700006BH, System ID NT24G1700006BH, Serial Number NT24G1700006
1100840682117869UDI/DI (01)00840682117869(11)17080021NT24G1700004BH, System ID NT24G1700004BH, Serial Number NT24G1700004
12UDI/DI Not Available, System ID Not Available, Serial Number NT24G1700008
13UDI/DI Not Available, System ID CS1052NM10, Serial Number T2CG15010/166BH9
14UDI/DI Not Available, System ID DK1043NM01, Serial Number T2CG15009/163BH6
15UDI/DI Not Available, System ID EG2559NM01, Serial Number NT24G1700001
16UDI/DI Not Available, System ID M4143529, Serial Number T2CG16018/187BH5
17UDI/DI Not Available, System ID 083026108252217, Serial Number T2CG16019/232BH9
18UDI/DI Not Available, System ID 083026108252217, Serial Number NT24G1700005
19UDI/DI Not Available, System ID 083026861420919, Serial Number NT24G1800007
20UDI/DI Not Available, System ID Not Available, Serial Number T2CG13006/127BH1
Amount recalled
116 units

This recall covers 13 products

The agency filed one notice per product. You are reading one of them; the other 12 are below.

Why it was recalled

GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Ge Medical Systems Israel, Functional Imaging recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:35 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.