Reported Jan 22, 2025 · Ongoing

GE Healthcare Discovery NM 630, Model/Catalog Numbers: 1) 5376204-30-2, 2) H2401CF, 3) H2401CH, 4) H2401JE, 5) H3100JY;…

Ge Medical Systems Israel, Functional Imaging

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING · 4, Hayozma St · Tirat Carmel · N/A

Is this the product you have?

  • GE Healthcare Discovery NM 630
  • Model/Catalog Numbers1) 5376204-30-2, 2) H2401CF, 3) H2401CH, 4) H2401JE, 5) H3100JY; System, Tomography, Computed, Emission
Lot codes and dates

525 of 525 products

Products covered by this recall
#BarcodesProduct as published
1UDI/DI N/A, System ID NT3058, Serial Number T2BG1600023/ 146187HL0
2UDI/DI N/A, System ID NT3061, Serial Number T2BG1600029/ 147600HL1
3UDI/DI N/A, System ID NT3111, Serial Number 147243HL0
4UDI/DI N/A, System ID NT3072, Serial Number 151822HL4
5UDI/DI N/A, System ID NT3081, Serial Number T2BG1700045/ 155928HL5
6UDI/DI N/A, System ID NT3080, Serial Number T2BG1700042/ 155272HL8
7UDI/DI N/A, System ID NT3098, Serial Number T2BG1700051/ 158839HL1
8UDI/DI N/A, System ID NT3088, Serial Number T2BG1700048/ 157050HL6
9UDI/DI N/A, System ID NT3073, Serial Number T2BG1700040/ 150067HL7
10UDI/DI N/A, System ID NT3076, Serial Number 153537HL6
11UDI/DI N/A, System ID NT3048, Serial Number 141262HL6
12UDI/DI N/A, System ID NT3083, Serial Number T2BG1700049/ 158144HL6
13UDI/DI N/A, System ID NT0003, Serial Number T2AG14033/ 108281HL7
14UDI/DI N/A, System ID NT3064, Serial Number T2BG1600025/ 146380HL1
15UDI/DI N/A, System ID NT3082, Serial Number T2BG1700047/ 156900HL3
16UDI/DI N/A, System ID NT3057, Serial Number T2BG1600022/ 146064HL1
17UDI/DI N/A, System ID NT0005, Serial Number T2AG14040/ 119081HL8
18UDI/DI N/A, System ID NT3054, Serial Number T2BG1600019/ 144873HL7
19UDI/DI N/A, System ID NT3070, Serial Number T2BG1600024/ 146363HL7
20UDI/DI N/A, System ID NT3067, Serial Number T2BG1600031/ 148143HL1
Amount recalled
704 units

This recall covers 13 products

The agency filed one notice per product. You are reading one of them; the other 12 are below.

Why it was recalled

GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Ge Medical Systems Israel, Functional Imaging recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 7:33 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.