Reported Oct 8, 2025 · Ongoing
MAGNETOM Vida (DE). Model Number: 11060815. 510(k): K170396, K173617, K181433, K183254, K192924, K203443, K213693,…
Siemens Medical Solutions USA, Inc.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355-1418
Is this the product you have?
- MAGNETOM Vida (DE)
- Model Number
11060815. 510(k): K170396K173617K181433K183254K192924K203443K213693K213805K220425K231560
- Lot codes and dates
- Model Number11060815
- UDI Numbers
498 of 498 codes
(01)04056869039176(21)176636(01)04056869039176(21)177211(01)04056869039176(21)175728(01)04056869039176(21)175685(01)04056869039176(21)175851(01)04056869039176(21)175768(01)04056869039176(21)177298(01)04056869039176(21)177295(01)04056869039176(21)176756(01)04056869039176(21)175754(01)04056869039176(21)177349(01)04056869039176(21)176757(01)04056869039176(21)177232(01)04056869039176(21)175935(01)04056869039176(21)176967(01)04056869039176(21)175741(01)04056869039176(21)177052(01)04056869039176(21)176060(01)04056869039176(21)176371(01)04056869039176(21)177012(01)04056869039176(21)177146(01)04056869039176(21)175770(01)04056869039176(21)176375(01)04056869039176(21)176122(01)04056869039176(21)176308(01)04056869039176(21)175669(01)04056869039176(21)176212(01)04056869039176(21)176755(01)04056869039176(21)176781(01)04056869039176(21)176302(01)04056869039176(21)176036(01)04056869039176(21)175766(01)04056869039176(21)176825(01)04056869039176(21)177151(01)04056869039176(21)176087(01)04056869039176(21)176771(01)04056869039176(21)177299(01)04056869039176(21)176451(01)04056869039176(21)177352(01)04056869039176(21)177354(01)04056869039176(21)176779(01)04056869039176(21)175870(01)04056869039176(21)175838(01)04056869039176(21)175692(01)04056869039176(21)177359(01)04056869039176(21)177119(01)04056869039176(21)175977(01)04056869039176(21)176037(01)04056869039176(21)177244(01)04056869039176(21)176686(01)04056869039176(21)175831(01)04056869039176(21)177096(01)04056869039176(21)176660(01)04056869039176(21)177034(01)04056869039176(21)176639(01)04056869039176(21)176397(01)04056869039176(21)175805(01)04056869039176(21)176358(01)04056869039176(21)176293(01)04056869039176(21)176118
- Serial Numbers
498 of 498 codes
176636177211175728175685175851175768177298177295176756175754177349176757177232175935176967175741177052176060176371177012177146175770176375176122176308175669176212176755176781176302176036175766176825177151176087176771177299176451177352177354176779175870175838175692177359177119175977176037177244176686175831177096176660177034176639176397175805176358176293176118
- Amount recalled
- 498 units
This recall covers 15 products
The agency filed one notice per product. You are reading one of them; the other 14 are below.
2 more are in the archive.
Where it was sold
- Nationwide
- GA
Why it was recalled
Stop using this product now
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Siemens Medical Solutions USA, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.