Reported Oct 8, 2025 · Ongoing
MAGNETOM Vida Fit. Model Number: 11410481.
Siemens Medical Solutions USA, Inc.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355-1418
Is this the product you have?
- MAGNETOM Vida Fit
- Model Number11410481
- Lot codes and dates
- Model Number11410481
- UDI Numbers
(01)04056869245188(21)202129(01)04056869245188(21)202025(01)04056869245188(21)202139(01)04056869245188(21)202016(01)04056869245188(21)202051(01)04056869245188(21)202055(01)04056869245188(21)202114(01)04056869245188(21)202119(01)04056869245188(21)202013(01)04056869245188(21)202050(01)04056869245188(21)202102(01)04056869245188(21)202160(01)04056869245188(21)202127(01)04056869245188(21)202071(01)04056869245188(21)202078(01)04056869245188(21)202123(01)04056869245188(21)202082(01)04056869245188(21)202107(01)04056869245188(21)202031(01)04056869245188(21)202110(01)04056869245188(21)202108(01)04056869245188(21)202101(01)04056869245188(21)202093(01)04056869245188(21)202104(01)04056869245188(21)202028(01)04056869245188(21)202026(01)04056869245188(21)202076(01)04056869245188(21)202142(01)04056869245188(21)202125(01)04056869245188(21)202155(01)04056869245188(21)202084(01)04056869245188(21)202151(01)04056869245188(21)202068(01)04056869245188(21)202135(01)04056869245188(21)202121(01)04056869245188(21)202006(01)04056869245188(21)202010(01)04056869245188(21)202011(01)04056869245188(21)202146(01)04056869245188(21)202145
- Serial Numbers
202129202025202139202016202051202055202114202119202013202050202102202160202127202071202078202123202082202107202031202110202108202101202093202104202028202026202076202142202125202155202084202151202068202135202121202006202010202011202146202145. Additional
- serial number added on 01/23/2026202135
- Amount recalled
- 41 units
This recall covers 15 products
The agency filed one notice per product. You are reading one of them; the other 14 are below.
2 more are in the archive.
Where it was sold
- Nationwide
- GA
Why it was recalled
Stop using this product now
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Siemens Medical Solutions USA, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.